Core body temperature of mice |
Accuracy of ±0.1°C |
Fast intermittent measurements |
Sensor Tip Diameter | .056'' |
Accuracy | ±0.1°C in the physiological range |
Time Constant: | 0.5 seconds |
Maximum Temperature: | 125°C |
Minimum Temperature | 0°C |
Lead Length | 5 foot |
Sterilization Protocol | Gas ETO, Gluteraldehyde, Isoprophyl Alcohol |
The RET-3ISO for mice has all the same features of the standard RET-3 plus:
• An isolated sensor (can be used with monitoring equipment where electrical interference from other electrodes may be present).
• Reinforced strain relief for applications where probe is in more frequent use.
• Smooth ball tip. Stainless steel shaft - 3/4'' long, .028'' diameter. Tip diameter .065'', 5 foot lead length. Time constant 0.5 seconds.
Time Constant is defined as the time required to reach 63% of final temperature in liquid. An accurate reading is obtained when the reading stops changing, typically 5-10 time constants.
Maximum Temperature refers to intermittent use of probe. When probe is used for continuous measurements, maximum temperature should be reduced by 25% unless specified.