Sensor Tip Diameter | 1/4'' |
Accuracy | 0.1°C in the physiological Range |
Time Constant: | 0.15 seconds |
Maximum Temperature: | 90°C |
Minimum Temperature | 0°C |
Lead Length | 5 feet |
Sterilization Protocol | Gas ETO, Gluteraldehyde, Isoprophyl Alcohol |
The SST-1 skin surface temperature probe for humans, to be taped on is re-useable and has a fast response time.
• Fast reading, with accuracy to 0.1°C .
• 1/4'' diameter 10Kt. gold sensor disc provides excellent skin contact. Tip diameter 1/4'', 5 foot lead length. Time constant 0.15 seconds. Not isolated
Time Constant is defined as the time required to reach 63% of final temperature in liquid. An accurate reading is obtained when the reading stops changing, typically 5-10 time constants.
Maximum Temperature refers to intermittent use of probe. When probe is used for continuous measurements, maximum temperature should be reduced by 25% unless specified.