Inexpensive, Accurate & Fast |
Flexible design - pack of 6 |
Sensor Tip Diameter | .029'' |
Accuracy | ±0.1°C in the physiological range |
Time Constant: | 0.01 seconds |
Maximum Temperature: | 350°C |
Minimum Temperature | -100°C |
Lead Length | 5 feet |
Sterilization Protocol | Gas ETO, Gluteraldehyde, Isoprophyl Alcohol |
PT-6 Sensor Six-Pack - For multi-point temperature measurements. Inexpensive, Flexible. Welded sensor beads, tough kapton insulation. Diameter is .029''. Maximum temperature 350°C, Time constant 0.01 Seconds. 5 foot lead. Not isolated.
Time Constant is defined as the time required to reach 63% of final temperature in liquid. An accurate reading is obtained when the reading stops changing, typically 5-10 time constants.
Maximum Temperature refers to intermittent use of probe. When probe is used for continuous measurements, maximum temperature should be reduced by 25% unless specified.