| Tissue Implantable/Research Only |
| Accurate to ±0.1°C |
| Small with Fast Response Time |
| Ideal for Animals Brains and Tissue |
| Sensor Tip Diameter | 0.011'' |
| Accuracy | ±0.1°C in the physiological range |
| Time Constant: | 0.005 seconds |
| Maximum Temperature | 150°C |
| Minimum Temperature | -100°C |
| Lead Length | 3 ft. |
| Sterilization Protocol | Autoclavable, ETO gas, Isoprophyl Alcohol and Gluteraldehyde solution |
IT-23 Flexible Implantable Microprobe - For ultra fast measurements and for use on micro-size specimens. Time constant 0.005 seconds. Tissue Implantable with a 23 gauge needle (supplied). Fairly fragile. Sensor wires are twisted TFE Teflon™ insulated. Maximum Temperature 150°C. Isolated. Probe diameter .011", 3 foot lead.
Time Constant is defined as the time required to reach 63% of final temperature in liquid. An accurate reading is obtained when the reading stops changing, typically 5-10 time constants.
Maximum Temperature refers to intermittent use of probe. When probe is used for continuous measurements, maximum temperature should be reduced by 25% unless specified.
PLEASE NOTE: WHEN USING FOR IMPLANTATION, the Physitemp IT-SERIES flexible probes are implantable temperature sensors designed for subcutaneous temperature measurement in research animals.Intended use: multiple general laboratory and calibration applications. Implantable and subcutaneous use is restricted to Animal Research Only. Not designed for implantation in humans or veterinary clinical practice applications.